Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the all-in-one-wp-security-and-firewall domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/rstechnics/domains/wp.rstechnics.nl/public_html/wp-includes/functions.php on line 6114
hand-held data logger – RS Technics BV

Regie-40

icon-support-regie40

icon-support-regie40

Support for Regie-40 hand-held data logger, portable recording system.
(no longer produced)


The printing of a graph constantly stops with error message.

In the event of a SENSOR-ERROR messages in graph (see the SE line under the measurement value graph) the printing of graphs is blocked.
This message can arise with:
a. Defective sensor or incorrectly connected.
b. No measurement signal is connected to the input.
First resolve these Sensor-Error messages through:

a. Replace defective sensors or connect properly
b. Set unconnected inputs to function OFF in window REGIE-40 – Change Parameters.

Then delete the Regie-40 and start up a new measurement period.
The data can now be printed out, if there are no longer any Sensor Error messages.
Comments: the data can however be requested in the Table display. 


Downloads

-installation package read out software v1.62NL

-read out/setting software manual v1.62NL